Herunterladen Inhalt Inhalt Diese Seite drucken

Magnescale Laserscale BL57-RE Anleitung Seite 42

Maßstabseinheit
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

• 0 V is the circuit ground, and it is not connected to the frame ground.
• Make sure that the power supply voltage is 5 V DC
• +5 V S and 0 V S are for checking the voltage (remote sensing function) applied to the input connector of
the interface unit. These voltages can be used to check and control for drops in the supply voltage due to
the cables. When using a power supply that cannot control power supply fluctuations, a power supply
input terminal can be used to reduce the supply voltage drops occurring due to the cable length. In this
case, connect the cable to the respective +5 V or 0 V power supply.
• The appropriate cable thickness is AWG28 to AWG24.
• Connect all of the 0 V terminals to prevent mis-wiring.
• Use shielded cables for all cabling. Connect the shield to the connector shell.
• Use twisted-pair cables for the output signals.
Use cables so that the following signals are paired:
A and
A, B and
B, REF and
ALM.
• For output format F or G, SIN (M) and COS (M) are dedicated signals for checking the scale signal level.
Never connect them to a device other than the oscilloscope. Also, use 0 V (M) for the GND.
6-2
(E)
±
5 % at the input connector to the interface unit.
REF, +SIN and −SIN, +COS and −COS, +REF and −REF, ALM and
BL57-RE

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis