4.2
Replacements
4.2.1
Replacing the Lamp
Unscrew the cap nut on the handpiece (item 14, foldout front page) by turning it counter-
•
clockwise.
Pull handpiece (item 14, foldout front page) from connection cable.
•
Insert new lamp in the lamp socket and put handpiece (item 14, foldout front page) back
•
on.
Unscrew the cap nut on the handpiece (item 14, foldout front page) by turning it
•
clockwise.
A damaged lamp does not have a negative effect on the general functioning of the
®
SonicWeld Rx
.
4.2.2
Replacing the Sonotrode
Unscrew sonotrode (item 19 – 22, foldout front page) from the handpiece (item 14, foldout
•
front page) by turning it counter-clockwise using a flat spanner, and tighten by turning it
clockwise with a maximum of 0.03 Nm.
42
Instructions for Use / Technical Description
SonicWeld Rx
®
Ultrasonic Unit
V. 2.2