Maintenance

4.1

Troubleshooting

Malfunction
Display Service after
pressing the mains
switch
"Screeching" sounds
during the self-test or
during the operation of
®
the SonicWeld Rx
.
Tip of sonotrode
cannot be removed
from the SonicPin Rx
No magnetic hold of
®
the SonicPins Rx
on
the sonotrode tip
Sonotrode tip "sticks"
®
on the SonicPin Rx
No light on the
handpiece
No display
Display ERROR during
operation
V. 2.2
Cause
Loose sonotrode
Components not
connected
Components defective
Residual moisture or
condensate on the
contact points.
®
Bent or rammed
sonotrode tip
Too long contact time
between sonotrode and
pin in regard to set
energy level
Bent sonotrode tip
Lamp damaged,
condensate on contact
points
No mains connection
Defective fuse
Device is not on
Device is overloaded
(jamming / jerking of
the sonotrode tip,
applied pressure too
high)
Detection
Display
Service
Clearly audible
Visual check
Pin is generally
deformed, sonotrode
cannot be removed
residue-free
Visual check
Visual check
Visual check
Electrical test
Mains switch is set to
"0"
Energy transfer is
interrupted
Correction
Tighten sonotrode using a
15
flat spanner (item
foldout front page)
Connect components
Replace defective
components
The sound does not affect
the proper functioning of the
®
SonicWeld Rx
. It will
usually stop by itself. Maybe
longer drying periods after
the sterilisation should be
considered.
Replace the sonotrode tip
Observe cooling time.
Loosen adhesion by turning
sonotrode in an axial way
(see chapter 2.4, page 36)
Replace the sonotrode tip
Consider replacing the lamp.
If the lamp on the sonotrode
is damaged, the ultrasonic
function will still work
properly
Connect to mains and turn
on device
Replace fuse
Set mains switch to "I"
Do not use the handpiece for
a few moments, and then
press foot switch / hand
activation again
,
41