Fault
software:
seca analytics 125
Value of an analysis parame-
ter is shown in red
007-276XX-XXX
013-276XX-XXX
to
018-276XX-XXX
020-276XX-XXX
021-276XX-XXX
030-276XX-XXX
034-276XX-XXX
036-276XX-XXX
037-276XX-XXX
039-276XX-XXX
Cause
Value outside the normal range deter-
mined for this analysis parameter
BIA board needs calibrating
Measurement invalid: Impaired by inter-
ference
BIA board defective
Invalid measurement
Impedance of BIA electrodes too high
Patient's hands and feet not positioned
correctly
Self-test failed
Temperature sensor defective
Ambient temperature too high or too
low
• Patient's hands and feet not posi-
tioned correctly
• Impedance of BIA electrodes too high
BIA electrodes defective
Remedy
• Repeat the measurement to rule out mea-
surement error
• If the value is still outside the normal
range, take this into account during anal-
ysis and further examinations
Inform seca Service
• Remove or switch off electronic devices
in the vicinity
• Repeat measurement
• If the error recurs, inform seca Service
Inform seca Service
• Repeat measurement
• If the error recurs, inform seca Service
• Moisten the patient's skin on the contact
surfaces (e.g. with electrode spray)
• If the error recurs, inform seca Service
• Make sure that the patient's hands and
feet are positioned correctly
• If the error recurs, inform seca Service
• If the error recurs, inform seca Service
Inform seca Service
Observe ambient conditions for operation,
transport, and storage
• Make sure that the patient's hands and
feet are positioned correctly
• Moisten the patient's skin on the contact
surfaces (e.g. with electrode spray)
Inform seca Service
Troubleshooting • 181