Fault/error code
099-297XX-XXX
10.4
Bioimpedance measurement
Fault
Foot silhouettes on the
weighing platform do not illu-
minate
is displayed during the
electrode check;
bioimpedance measurement
does not start
is displayed continuously;
bioimpedance measurement
does not start
is displayed
software:
seca analytics 125
Results of bioimpedance
measurement deviate signifi-
cantly from expected results
180 • seca 555/554/552; seca 550/549; seca 257/256
Cause
Voice guidance: Language that does
not support the announcement of mea-
sured results in imperial units is active,
announcement of measured results de-
activated automatically
Cause
Device is in an undefined state follow-
ing implausible input
Foot silhouette illumination is defective
Patient's hands or feet not positioned
correctly
Patient's skin too dry
Patient's skin too calloused
Electrodes defective
Bioimpedance measurement failed
Patient moved during the measurement
Patient used different pairs of hand
electrodes on the left and right
Electrodes defective
Remedy
è Switching
• Set metric units
units, page 160
• Activate announcement of measured re-
è Activating/deactivating an-
sults
nouncement of measured results (Re-
sults), page 171
• Select a language that does support the
announcement of measured results in im-
perial units: EN-US, EN-UK, ES-MX, ES-
è Selecting language, page 168
SP
• Activate announcement of measured re-
è Activating/deactivating an-
sults
nouncement of measured results (Re-
sults), page 171
Remedy
è Restarting the device, page 184
• If the error recurs, inform seca Service
Inform seca Service
Make sure that the patient's hands and
feet are positioned correctly:
• Feet on the illuminated foot silhouettes
• Hands on the same hand electrodes on
both sides
Spray the skin with a little electrode spray
at the contact points
Spray the skin with a little electrode spray
at the contact points
Inform seca Service
• Clear measurement with
• Check patient's electrode contact
• Repeat bioimpedance measurement
Ask the patient not to move during the
measurement and repeat the measurement
Ensure that the patient uses the same
hand electrodes on both sides and repeat
the measurement
Inform seca Service