10.5
Data transmission
Fault/error code
The "Date of birth" dialog
window appears after the
key is pressed
After the
key is pressed,
there is no request for patient
or user data
icon appears
icon appears
icon appears
key appears grayed out
182 • seca 555/554/552; seca 550/549; seca 257/256
Cause
Patient ID not scanned
ID scanning not set up
Patient does not yet have a file in the
EMR System
Device is connected to the
software
seca analytics 125
ID(s) scanned at the start of the mea-
surement
ID scanning not set up
function activated
Anonymous
An obligatory measurement has not
been performed
Scanned ID is invalid
Optional measured value (e.g. height)
has not been determined
Optional ID (e.g. user ID) has been
scanned
There are no data available to be con-
firmed
Remedy
• Press the
key
• Scan patient ID
Check Workflow settings for the
or
seca connect 103
seca analytics 125
software
• Enter date of birth and press
again
• Create patient file in the EMR System and
assign measurement
• Enter date of birth
• Press the
key
• Not a malfunction, measured results are
assigned to the patient and saved in the
EMR System
• Check assignment in EMR System
Check Workflow settings for the
seca connect 103
or
seca analytics 125
software
• Not a malfunction, measured results are
sent to the EMR System
• In the EMR System, ensure that mea-
sured results are correctly assigned
• Press the
key
• Perform the missing measurement
Scan valid ID
• Press
key and determine optional
measured value
• Press
key and end measurement
• If desired: Check Workflow settings for
the
seca connect 103
seca analytics 125
software
If desired: Check Workflow settings for the
or
seca connect 103
seca analytics 125
software
• Perform a measurement
• Scan IDs (user/patient)
key
or