Herunterladen Diese Seite drucken

Seca mBCA 555 Bedienungsanleitung Seite 197

Vorschau ausblenden Andere Handbücher für mBCA 555:

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Measured values
Phase angle measuring range
Impedance measuring range
Measuring segments
Measuring current
Measuring time
Accuracy (frequencies: 1; 2; 5; 10; 20; 50 kHz, segments: right half of
body, left half of body):
• Impedance (phase angle 0°)
• Phase angle (phase angle 0°), impedance 200 Ω to 1000 Ω)
Minimum age of patient
Height of patient
Analysis parameters
13
OPTIONAL ACCESSORIES AND SPARE PARTS
Accessory/spare part
Switch-mode power supply:
100-240 V~ / 50-60 Hz, 12 V= / 1.5 A / 18 W
Barcode scanner
holder for barcode scanner
seca 463
seca 459
"panda" figure
"animals" sticker for ultrasonic measuring rods
seca 487
Handrails
Measuring rods
Configuration software
Analysis software
Technical data, bioimpedance measurement
Impedance (Z), resistance (R),
reactance (X
Right arm, left arm, right leg, left leg, right
half of body, left half of body, torso
100 μA (+20 %, -50 %)
No result display on the device, see in-
structions for use for
Article number
See recommendation at www.seca.com
è Compatible seca products, page 198
è Compatible seca products, page 198
Optional accessories and spare parts • 197
English
), phase angle (φ)
c
0° to 20°
10 Ω to 1000 Ω
max. 30 s
± 5 Ω
± 0.5°
5 years
≥ 130 cm
seca analytics 125
software
68 32 10 273
463 0000 009
459 0000 009
487 0045 009

Werbung

loading

Diese Anleitung auch für:

Mbca 554Mbca 552Mbca 550Mbca 549257256