5.3
Setting up device
To achieve accurate measured results, the floor at the setup location must be
level and stable. Soft floors (wooden boards, for example) give under the pa-
tient's weight and falsify the measured result.
1. Place the device on a firm, level surface.
2. Only for devices with an ultrasonic measuring rod: Mark the area shown in
the illustration using colored adhesive tape, for example.
min
min. 0,5 m
min
Example device configuration: seca 555 scale
seca 550 BIA handrail, seca 257 ultrasonic measuring rod
NOTICE!
Malfunction caused by other ultrasonic emitters
If there are other ultrasonic emitters in the immediate vicinity of the de-
vice - automatic door openers, for example - incorrect measurements
will result.
► Ensure that there are no other ultrasonic emitters in the same room
or in the immediate vicinity of the device.
min. 0,2 m
min. 0,2 m
min. 0,2 m
min. 0,2 m
,2
,2
m
m
min. 0,5 m
min. 0,5 m
min. 0,5 m
Starting up device • 125