Handling packaging material
NOTICE!
Inconsistent measuring results
► Before you save and continue using measured values determined
using this device (e.g. in seca software or in an EMR system), en-
sure that the measured values are plausible.
► If measured values have been sent to seca software or to an EMR
system, ensure before continuing to use them that the measured
values are plausible and assigned to the correct patient.
NOTICE!
Measuring results from other devices not compatible
Bioimpedance measurements performed by devices from different
manufacturers are not compatible. Follow-up measurements not per-
formed on a seca device may lead to inconsistent data and to misin-
terpreted measuring results.
► Ensure that follow-up measurements are also performed on a seca
device.
NOTICE!
Malfunction caused by other ultrasonic emitters
If there are other ultrasonic emitters in the immediate vicinity of the de-
vice - automatic door openers, for example - incorrect measurements
will result.
► Ensure that there are no other ultrasonic emitters in the same room
or in the immediate vicinity of the device.
NOTICE!
Faulty measurement due to reflections
If there are objects or people in the immediate vicinity of the device, in-
correct measurements will result.
► Ensure that there are no objects or people within 0.5 meters of the
front or side of the scale during the measurement procedure.
► Ensure that the device is at least 0.2 meters away from the wall.
► Ensure that the patient is not wearing any kind of hair accessory on
top of their head.
WARNING!
Risk of suffocation
Packaging materials made of plastic film (bags) present a risk of suffo-
cation.
► Store packaging material out of the reach of children.
► If the original packaging material is no longer available, only use
plastic bags with safety holes to reduce the risk of suffocation. Use
recyclable materials if possible.
NOTE
Store the original packaging material for future use (e.g. returning for
servicing).
English