Herunterladen Diese Seite drucken

Seca mBCA 555 Bedienungsanleitung Seite 183

Vorschau ausblenden Andere Handbücher für mBCA 555:

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Fault/error code
key appears grayed out
Autoclear function: An entry
of "0 seconds" is not ac-
cepted; the factory setting
(300 seconds) is suggested
instead
Workflow LED does not light
up
Workflow LED illuminated red
001-288XX-XXX
002-288XX-XXX
004-288XX-XXX
005-288XX-XXX
006-288XX-XXX
Cause
There are no data available to be
cleared
Implausible input; switching off the
function is not intended in the factory
Measurement not yet started
No network connection
WiFi function deactivated
Workflow LED defective
ID not found in EMR System or in the
seca software
Device has not saved measured results
to the clipboard
Measured results have not been sent to
an EMR System or to seca software
No connection to the server
No data transmission possible
Barcode scanned at the wrong time
Barcode invalid or damaged
Patient ID/user ID not found
Unable to transmit measurement to
EMR System
Remedy
• Perform a measurement
• Scan IDs (user/patient)
• Accept factory setting
• Manually enter value between 1 and 3600
seconds
Start measurement
(è Starting the mea-
surement
procedure, page 130)
Set up network connection
network functions, page 161
Activate the WiFi function
activating the WiFi function, page 163
Inform seca Service
Create the ID in the EMR System or in the
seca software
Repeat measurement
• Repeat measurement
• Check WiFi connection
• Check LAN cable
• Check network settings
network functions, page 161
• Check network settings
network functions, page 161
• Check Workflow settings for the
seca connect 103
or
seca analytics 125
software
Repeat measurement and maintain correct
operating sequence
è Restarting the device, page 184
• If the error recurs, inform seca Service
Use a valid, undamaged barcode
In the database (
seca analytics 125
EMR System):
• Create patient file
• Create user account
Check Workflow settings for the
seca connect 103
or
seca analytics 125
software
Troubleshooting • 183
English
è Setting up
è Activating/de-
è Setting up
è Setting up
or

Werbung

loading

Diese Anleitung auch für:

Mbca 554Mbca 552Mbca 550Mbca 549257256