5.6
Installing the drip guard
128 • seca 555/554/552; seca 550/549; seca 257/256
1. Plug the adapter required for your power supply into the power supply unit.
2. Insert the device connector of the power supply unit into the power supply
connection socket of the device.
3. Plug the power supply unit into a power supply socket.
4. Perform a function check
NOTICE!
Damage to device due to the ingress of fluids
Damage may occur to the device if dripping water or other dripping
fluids enter the device via the connection panel.
► Only operate the device with the drip guard installed.
1. Connect all cables as described in these instructions for use and in the in-
stallation instructions for the compatible products.
2. Position the drip guard on the connection panel as shown in the illustration.
3. Push the drip guard beneath the weighing platform up to the stop.
è Function
check, page 176.