8. Place the calibration rod centrally on the illuminated foot silhouettes of the
weighing platform.
9. Step back from the measuring device (distance approx. 0.5 m).
ð The second step of the calibration procedure starts.
10. Wait until the second part of the calibration procedure has been completed.
ð The main screen is displayed again.
11. Remove the calibration rod from the weighing platform.
ð The device is ready to measure.
English
Configuration • 153