Herunterladen Diese Seite drucken

Seca mBCA 555 Bedienungsanleitung Seite 177

Vorschau ausblenden Andere Handbücher für mBCA 555:

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Fault
Multifunctional display does
not react when keys are
pressed
Multifunctional display re-
mains dark after pressing
10.2
Measuring weight
Fault/error code
Displayed weight is implausi-
ble
Autohold
function cannot be
deactivated
001-272XX-XXX
to
008-272XX-XXX
010-272XX-XXX
013-272XX-XXX
016-272XX-XXX
019-272XX-XXX
Cause
Device is in an undefined state follow-
ing implausible input
With combinations of devices involving
a handrail and/or measuring rod: Addi-
tional display connected
No connection to weighing platform
No plug-in power supply unit con-
nected
Device is in an undefined state follow-
ing implausible input
Multifunctional display defective
Cause
Weighing electronics using outdated
zero point
Weighing electronics defective
Device is connected to a network (in-
tended behavior):
Autohold
activated automatically
Load cell or weight calculation module
defective
Scale has been switched on with too
high a load
Scale was caused to oscillate and was
unable to determine the zero point
Maximum capacity exceeded
Ambient temperature too high or too
low
è Restarting the device, page 184
• If the error recurs, inform seca Service
è Restarting the device, page 184
• If the error recurs, inform seca Service
• Check whether the display cable is con-
nected correctly
• If the error recurs, inform seca Service
• Check whether the plug-in power supply
unit is connected correctly
• If the error recurs, inform seca Service
è Restarting the device, page 184
• If the error recurs, inform seca Service
Inform seca Service
• Remove the weight from the weighing
platform
• Press
Weight
è Restarting the device, page 184
• Wait until main screen is displayed again
Inform seca Service
If necessary, disconnect device from net-
function is
work
Inform seca Service
è Restarting the device, page 184
• If the error recurs, inform seca Service
è Restarting the device, page 184
• If the error recurs, inform seca Service
Ask patient to step off the scale
• Observe ambient conditions for opera-
tion, transport, and storage
technical data, page 193
English
Remedy
Remedy
display field
è General
Troubleshooting • 177

Werbung

loading

Diese Anleitung auch für:

Mbca 554Mbca 552Mbca 550Mbca 549257256