Permitting anonymous measurements
166 • seca 555/554/552; seca 550/549; seca 257/256
Device mode
Basic
Advanced
Expert
Service
If the device is connected to a suitable EMR System via the
software, you can permit anonymous measurements. If you activate this func-
tion, the device does not request user ID or patient data (date of birth, patient
ID).
WARNING!
Incorrect assignment of measured results, data loss
If measured results are incorrectly assigned or lost, this will lead to
misinterpretations and consequently to misdiagnoses.
► Ensure that your work environment supports anonymous measure-
ments so that clear assignment of measured results is always as-
sured.
► Use this function only in consultation with seca Service.
1. Press the
key.
ð The
Settings
menu is displayed.
2. Press the
or
key until the
3. Press the
item.
Network
ð The current setting is displayed.
4. Press the desired setting for the
Function activated
Function deactivated
5. To exit the menu, press the
Function available
seca connect 103
Network
menu item is displayed.
item:
Anonymous
key.