Error Messages

Error Messages

Error Messages and their Reasons

Table 4
Error Message
Probable Cause
Repositioning the
Error Moving
zero plate is not
Zero Plate
possible due to a
bubble in sample
or reference cell.
Dirty measurement
cell
Repositioning the
zero plate is not
possible due to a
malfunction.
Signal Overflow Excessive light
intensity in sample
or reference cell
Error Messages
Solution
Flush sample and reference cell
several times. Please use for
example methanol or water
methanol mixtures instead of
pure water. Check I
and I
1
follow the instructions in
‚AUTOZERO ' on page 15.
Please clean the cell.
Please contact KNAUER
service depatment.
Please flush sample and
reference cell again and press
the AUTOZERO key.