Herunterladen Inhalt Inhalt Diese Seite drucken

Technical Data - Knauer Smartline 2300 Benutzerhandbuch

Ri detektor 2300
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Technical Data

Technical Data
Measurement angle
Refractive index
(range)
Cell volume
Measurement range
Sensitivity
Noise
Wavelength
Capillary ID
inlet:
sample cell:
reference cell:
outlet:
Maximum flow rate
Time constant
Scalable integrator
output
Autozero range
Display
Control
Special features
Power supply
Dimensions
(W x H x D)
Weight
GLP-report
Smartline 2300
(analytical)
45°
1.00 – 1.75 RIU
15 µl
-3
± 1 x 10
RIU
-8
3 x 10
RIU
-8
≤ ± 1.5 x 10
RIU
950 ± 30 nm
0.3 mm
1.0 mm
1.0 mm
0.7 mm
5 ml/min
0.1 0.2 0.5 1.0
2.0 5.0 und 10 s
+ 1.0 V ,
16 steps
Full range
LCD,
2 x 24 digits
RS 232 interface,
analog output,
remote
connector,
keypad
Integrated flush valve
for rinsing the
reference cell
90-260 V, 47-63 Hz,
30 VA
226x135x390mm
8 kg
Operating hours,
serial number
33
Smartline 2400
(preparative)
15°
1.00 – 1.75 RIU
9 µl
-3
± 2 x 10
RIU
-7
4 x 10
RIU
-7
≤ ± 2 x 10
RIU
950 ± 30 nm
1.0 mm
1.0 mm
1.0 mm
1.0 mm
100 ml/min
0.1 0.2 0.5 1.0
2.0 5.0 und 10 s
+ 1.0 V ,
16 steps
Full range
LCD,
2 x 24 digits
RS 232 interface,
analog output,
remote
connector,
keypad
Integrated flush valve
for rinsing the
reference cell
90-260 V, 47-63 Hz,
30 VA
226x135x390mm
8 kg
Operating hours,
serial number

Werbung