26
Fig. 19
Installing the Measurement Cell
SOP 5
Capillaries
Site of Measurement Cell in the RI Detector 2400
You can clean the measurement cell in an ultrasonic bath filled with a
suitable cleaning solution, or install a new one.
Make sure that all parts are properly functioning before installing.
It is recommended to use a new set of seals whenever installing a
cell.
Installing the Measurement Cell of RI Detector 2300
1. Put the lower seals over the capillaries' ends in the cell ground,
taking the bevels into consideration.
2. Insert the measurement cell, considering it's position (see ) into the
housing. The writing KNAUER has to be situated on the upper side.
Measurement
cell unit
Simple Maintenance
Optical
bench