Herunterladen Inhalt Inhalt Diese Seite drucken

Knauer Smartline Pump 1000 Handbuch Seite 37

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

36
Table 5
Connecting the Smartline Pump 1000 to other Instruments
Communication example for the 10 ml pump head
Input
Reply
ST 2
OK
ST 2.200
OK
ST 10
?
Other commands permissible via RS-232 interface to the Smartline pump
1000:
Serial number inquiry
SN
GLP inquiry
GLP
date and time inquiry
Date
Program version inquiry
VER
Start the pump, using the currently selected flow rate
ST
Stops the pump
SP
minmal pressure inquiry
Pmin
maximal pressure inquiry
Pmax
Pressure inquiry in Mpa
P
CONTROL
Enables control from the key pad
STANDALONE
CONTROL
Enables control from the serial port
REMOTE
If using serial control, the pump sends the following messages to the PC:
The last command was successfully executed
OK
The last command could not be understood and/or
E:command
executed.
Description
Set the flow rate to 2 ml/min Command
correctly understood and executed
Set the flow rate to 2.200 ml/min
Command correctly understood and
executed
Set the flow rate to 10 ml/min Command
not understood. Flow rate remains at
2.200 µl/min