Connecting the Smartline Pump 1000 to other Instruments
Table 3
Fig. 27
Table 4
Connecting the EVENTS socket
You can control the states of individual contacts of the EVENTS socket
using SOP 9 "Setting the Control Signals (Events" on page 19.
Specifications of the Event signals
TTL (Transistor Transistor Logic) connections are to be connected only to
TTL compatible outputs (max. consumption: 40 mA, U
directed to GROUND.
Technical specifications of the Event signals
State
Control outputs
0
Low (< 500 mV)
1
High (> 2 V)
Pulse,
2
> 2 V for 500 ms
Connecting the Analog/Error/Start Socket
The Analog/Error/Start terminal strip serves to send/receive either start or
error signals from/to other instruments. An additional voltage signal is
available for monitoring the system pressure or for controlling a slave
pump (see item ANALOG OUTPUT in "Using the SETUP Menu" on page
13). The ANALOG IN enables the external analog control of the pump.
The REMOTE terminal strip
Technical specifications and description of Analog/Error/Start signal
connections
Signal
Start connections
START IN
START OUT
Error- connections
ERROR IN
ERROR OUT
Analog connections
ANALOG OUT
ANALOG IN
OC (Open Collector)-
connections
Inactive
Active
Pulse,
active for 500 ms
Description
Activated by a 0 Volt or low signal or a short
circuit; I
= 10 mA
IN
OC (open collector) output is active for 500 ms
When receiving a 0 V (error) signal, the
message Error signal was detected is
displayed and the pump is stopped; I
OC output is active until the error condition is
removed
Analog voltage signal gives either measured
system pressure (1 V per 40 MPa; Offset for 0
MPa is some mV) or a control voltage for slave
pump B. Adjustable voltage range between 1,
2, 5, and 10V maximum output.
Analog input signal controls flow rate; default 1
V per 1 ml/min for 10 ml head and 5 ml/min for
33
= 24 V)
CEmax
Relay connections
Contact open
Contact closed
Pulse: Contact closed
for 500 ms
= 10 mA
IN