19.0 Apply Strain Relief Boot
1. The strain relief boot (found in the Surgical Connector Kit, part 22-0028) is pre-
slit along the center of one of the flat surfaces to aid placement over the
electrode wires.
2. Gently open the SLIT on the strain relief boot and wrap around the electrode
leads and then slide the boot over the electrode connector block.
3. Turn the connector over (opposite of SLIT side) and hold the front flap of the boot
up and remove the adhesive liner with forceps. Press the boot down firmly to the
connector as shown.
31