2. Lay silver pin end of electrode into needle tip bevel area.
3. Keep electrode parallel to needle to not damage silicone sheath on
electrode while loading through the tube of the tool.
4. DO NOT damage electrode by forcing through the needle bevel.
5. Continue to push electrode through the needle until the tip of the
electrode nears the bevel of the needle.
6. Place thumb against bevel of the needle point and pull the remaining
electrode through the tube of the tool. Ensure the blue tip is
completely in the bevel of the needle.
7. Verify that only the electrode stimulating tip of the electrode is exiting the
needle tip of the Disposable Electrode Delivery Tool.
8. While holding handle, unlock the needle position by pushing the slide
button.
·
Option #2: Loading electrode with needle at 90̊ :
1. Squeeze handle to the first locking position and lock the needle by pushing
the slide button.
2. Lay silver pin end of electrode into needle tip bevel area.
22