Appendix A: Electromagnetic Emissions and Immunity
Guidance and manufacturer's declaration - electromagnetic immunity
®
Propex IQ
Apex Locator is intended for use in the electromagnetic environment specified below.
The customer or the user of Propex IQ
environment.
Immunity test
IEC 60601 test level
standard
Electrostatic
discharge
± 6 kV contact
(ESD) IEC
± 8 kV air
61000-4-2
± 2 kV for power
Electrical fast
supply lines
transient / burst
± 1 kV for input /
IEC 61000-4-4
output lines
± 1 kV line(s) to
Surge IEC
Line(s)mode
61000-4-5
± 2 kV line(s) to earth
< 5% UT
(> 95% dip in UT) for
0.5 cycle
< 40% UT
Voltage dips,
(60% dip in UT) for 5
short
cycles
interruptions
and voltage
variations on
70% UT
power supply
(30% dip in UT) for 25
lines IEC
cycles.
61000-4-11
< 5% UT
(> 95% dip in UT) for
5 s
END-IFU-PROPEX-IQ-V06-WEB-EUR-Multilingual-2021-01
®
Apex Locator should ensure that it is used in such an
Compliance level
± 8 kV contact
± 15 kV air
± 2 kV for power
supply lines
Not applicable (no
input/output lines)
± 1 kV differential
mode
Not applicable (no
earth)
0% UT
(> 95% dip in UT) for
0.5 cycle
(at 0, 45, 90, 135,
180, 225 and 315°
phase angle)
0% UT
(> 95% dip in UT) for 1
cycle
(at 0 phase angle)
70% UT
(30% dip in UT) for 25/
30 cycles (at 0 phase
angle)
0% UT
(> 95% dip in UT) for
250/300 cycles (at 0
phase angle)
Electromagnetic environment
- guidance
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.
65/74