Troubleshooting
Apical Reverse function cannot be switched on.
Possible causes
No connection between Propex IQ
Locator and X-Smart IQ
Wrong cable connection.
Cable or connector damaged.
®
X-Smart IQ
Handpiece not connected to
®
Endo IQ
App.
Firmware of X-Smart IQ
up to date.
Apical Reverse setting on the Endo IQ
App accidentally set to OFF.
Electronic malfunction.
®
Propex IQ
Apex Locator is overheating.
Possible causes
Electronic malfunction.
®
Propex IQ
Apex Locator has overheated
and it has switched off.
END-IFU-PROPEX-IQ-V06-WEB-EUR-Multilingual-2021-01
®
Apex
®
Handpiece.
®
Handpiece not
®
Solutions
Ensure proper connection of the cables
between Propex IQ
®
the X-Smart IQ
Handpiece.
Ensure that all cables are original and
undamaged.
Connect X-Smart IQ
benefit from Apical Reverse function.
Ensure that X-Smart IQ
firmware is the latest version available
(updated for Propex IQ
feature).
Switch ON the feature on the Endo IQ
App and retry.
Perform a factory reset.
Contact Dentsply Sirona Endodontics
Customer Service.
Solutions
Allow the device to cool down and start
again.
Check for possible leakage from
®
Propex IQ
Apex Locator (liquids from
defective battery). In case of suspected
leakage, refer to Section
®
Apex Locator and
®
Handpiece to app to
®
Handpiece
®
Apex Locator
®
3