Herunterladen Inhalt Inhalt Diese Seite drucken

Dentsply Sirona Propex IQ Gebrauchsanweisung Seite 58

Apex-lokalisator
Vorschau ausblenden Andere Handbücher für Propex IQ:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 225
Troubleshooting
Battery does not recharge when the power supply is connected
(the battery LED remains OFF when power supply is connected).
Possible causes
Bad charger connection.
Power socket is not under tension/
switched off (if applicable).
Charger malfunction.
Electronic malfunction.
Device test and quick closed-loop test do not start.
Possible causes
®
Propex IQ
External tester is not working
properly.
Electronic malfunction.
®
X-Smart IQ
Handpiece undetected in combined mode.
Possible causes
®
X-Smart IQ
Measurement Cable or one
of the connectors are damaged.
Firmware of the X-Smart IQ
not up to date.
Electronic malfunction.
58/74
Solutions
Solutions
Solutions
®
Handpiece
END-IFU-PROPEX-IQ-V06-WEB-EUR-Multilingual-2021-01
Disconnect the charger from the device
and reconnect it again.
Replace the charger with an original
charger.
Contact Dentsply Sirona Endodontics
Customer Service.
Perform a factory reset.
Contact Dentsply Sirona Endodontics
Customer Service.
Check the connection of the cable
®
between Propex IQ
®
the X-Smart IQ
Handpiece.
Perform a closed- loop test in combined
mode.
®
Check if X-Smart IQ
is the latest version available (updated for
®
Propex IQ
Apex Locator feature).
Perform a factory reset.
Contact Dentsply Sirona Endodontics
Customer Service.
Apex Locator and
Handpiece firmware

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis