Troubleshooting
No feedback from Propex IQ
Possible causes
•
The battery is completely empty (all LEDs
are OFF).
•
The cables are damaged or not connected
correctly.
•
Non-original cables are used.
•
If all else fails.
END-IFU-PROPEX-IQ-V06-WEB-EUR-Multilingual-2021-01
®
Apex Locator during treatment.
Solutions
•
•
•
•
•
•
•
•
•
Connect the device to the charger and
charge the device.
Ensure that original cables have been
used.
Ensure that the cables are undamaged
and connected correctly.
Perform closed-loop to test to check
functionality of cables and device test to
check functionality of device.
If in combined mode, ensure that the
®
original X-Smart IQ
being used.
If in combined mode, ensure that the
endodontic file has been correctly inserted
®
into the X-Smart IQ
Clean the lip clip and the file clip according
to the procedure.
Perform a factory reset.
If cables are damaged or if the
Closed-Loop test fails after a
second attempt, DO NOT USE
®
Propex IQ
Apex Locator with the
damaged cables and contact your
Dentsply Sirona Endodontics
Customer Service.
If requested to send the device and/or
components to Dentsply Sirona
Endodontics Customer Service, reprocess
the device, file clip, lip clip and cables
before sending them back to customer
services (Refer to Section
Disinfection and
Sterilization).
Contra-angle is
Contra-angle.