Troubleshooting
Endodontic file progression indication is unstable or frozen.
Possible causes
LEDs damaged or not working properly.
Electronic malfunction.
Endodontic file progression indication is too sensitive (i.e. it displays the relevant position
for working length determination too quickly).
Possible causes
Short-circuit due to excess fluid in the pulp
chamber (rinsing solution, saliva, blood).
Direct contact between endodontic file and
gingiva / gingival proliferation (i.e. in cases
with fractured metal crowns).
Direct contact between endodontic file and
metal restoration work (crowns, parapulpal
pins; amalgam fillings).
Lateral root canal.
Immature canal with large apex.
END-IFU-PROPEX-IQ-V06-WEB-EUR-Multilingual-2021-01
Solutions
Perform a factory reset.
Contact Dentsply Sirona Endodontics
Customer Service.
Solutions
Dry the access cavity using an air blower
or a cotton ball. In case of excess bleeding
wait until it has stopped.
If heavy bleeding is observed, wait until
bleeding has been successfully stopped.
For isolation:
- Adequate tooth build-up;
- Crown lengthening;
- Electro-cauterisation.
Use a dental dam. Isolate the endodontic
file by placing 2-3 silicone stoppers on it or
insert the endodontic file in a small
polyvinyl tube before use.
Use the protective X-Smart IQ
Handpiece Sleeve on the X-Smart IQ
Handpiece (if in combined mode).
The use of X-Smart IQ
Handpiece Sleeve on the
X-Smart IQ
working in combined mode) is
mandatory to ensure the
correct functioning of the
system in combined mode.
Carefully widen the access cavity. If
necessary add flow composites to isolate.
Repeat length determination process.
®
®
®
®