E
Calibration is complete
Unplug power plug of vacuum pump from socket on backside of base unit and plug it into
automatic module (original state)
Unplug power plug of automatic module and plug it into socket on the backside of the base unit
(original state)

21. Auto Start Stop

With Auto Start Stop fully automatic distillation is started or stopped. Before such action, enter all
distillation parameters and start the following functions with the Auto Start Stop key:
Lift UP
Rotation ON
Heater ON
Vacuum pump and vacuum controller is activated
Product feed and condensate discharge is activated
Start timer (LED Act Time flashing)
An LED next to the Auto Start Stop key confirms such action.
CAUTION: if one or more functions are already active, e.g. heater, only the remaining functions will
be started.
The Time display shows time elapsed from pressing the Auto Start Stop key.
Pressing the Auto Start Stop key once more will reset the functions started:
Lift DOWN
Rotation OFF
Heater OFF
Aerating valve OPEN, vacuum control OFF and vacuum pump OFF
Product feed and condensate discharge is cancelled
During this automatic mode, all distillation parameters may be changed!