Reaching this final pressure during distillation, will discontinue distillation right away. Also refer to
1 2 5
EndP
Auto Start Stop function (see page 94)
If distillation should not be discontinued, enter "1" in "Endp mode", which never can
be reached by physical condition.
For hysteresis setting select LED

20. Calibrate Vacuum pressure sensor

Install separate calibrated vacuum gauge in vacuum
line close to condenser
Unplug power plug of automatic module from socket
of base unit
Unplug power plug of vacuum pump from socket of
the backside of automatic module and plug into plug
on the backside of base unit
All manuals and user guides at all-guides.com
Display I
Display II
Vapor. Use Set keys as required
Taste Select
92
Taste Start Stop