• Portable RF communications equipment, including peripherals such as antenna cables and external antennas, should
be no closer than 12 inches (30 cm) to any part of P P r r o o C C u u i i t t y y bed series, including cables specified by the manufacturer.
• Avoid stacking or placing equipment adjacent with other equipment to prevent improper operation of the product. If such
use is necessary, carefully observe stacked or adjacent equipment to make sure that they operate properly.
• The use of accessories, transducers, and cables, other than those specified or provided by the manufacturer, could
result in increased electromagnetic emissions or decreased electromagnetic immunity and result in improper operation.
The 3009 P P r r o o C C u u i i t t y y bed series was evaluated using the following cables:
AC mains input cable
AC aux input cable
AC aux output cable
Nurse call (DB-37)
G G u u i i d d a a n n c c e e a a n n d d m m a a n n u u f f a a c c t t u u r r e e r r ' ' s s d d e e c c l l a a r r a a t t i i o o n n - - e e l l e e c c t t r r o o m m a a g g n n e e t t i i c c e e m m i i s s s s i i o o n n s s
The 3009 P P r r o o C C u u i i t t y y bed series is intended for use in the electromagnetic environment specified below. The customer or
the user of the 3009 P P r r o o C C u u i i t t y y bed series should assure that it is used in such an environment.
E E m m i i s s s s i i o o n n s s t t e e s s t t
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations
Flicker Emissions
IEC 61000-3-3
The 3009 P P r r o o C C u u i i t t y y bed series is suitable for use in a professional healthcare facility environment and not in
environments exceeding immunity test conditions that the product was evaluated to, such as near high frequency (HF)
surgical equipment and inside of the radio frequency (RF) shielded room of magnetic resonance imaging (MRI)
equipment. The customer or the user of the 3009 P P r r o o C C u u i i t t y y bed series should assure that it is used in such an
environment and that the electromagnetic environment guidance listed below is followed.
I I m m m m u u n n i i t t y y t t e e s s t t
Electrostatic Discharge
(ESD)
IEC 61000-4-2
EN
48
C C a a b b l l e e
USB cable
Pendant
C C o o m m p p l l i i a a n n c c e e
Group 1
Class A
Class A
Complies
G G u u i i d d a a n n c c e e a a n n d d m m a a n n u u f f a a c c t t u u r r e e r r ' ' s s d d e e c c l l a a r r a a t t i i o o n n - - e e l l e e c c t t r r o o m m a a g g n n e e t t i i c c i i m m m m u u n n i i t t y y
I I E E C C 6 6 0 0 6 6 0 0 1 1 t t e e s s t t l l e e v v e e l l
±8 kV contact
±15 kV air
L L e e n n g g t t h h ( ( m m ) )
2.5
2.5
2.2
4.5
4.7
5.3
E E l l e e c c t t r r o o m m a a g g n n e e t t i i c c e e n n v v i i r r o o n n m m e e n n t t
N N o o t t e e - - The emissions characteristics of this
equipment make it suitable for use in industrial areas
and hospitals (CISPR 11 class A). If it is used in a
residential environment (for which CISPR 11 class B
is normally required) this equipment might not offer
adequate protection to radio-frequency
communication services. The user might need to take
mitigation measures, such as relocating or re-
orienting the equipment.
C C o o m m p p l l i i a a n n c c e e l l e e v v e e l l
±8 kV contact
±15 kV air
E E l l e e c c t t r r o o m m a a g g n n e e t t i i c c
e e n n v v i i r r o o n n m m e e n n t t - - g g u u i i d d a a n n c c e e
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
3009-009-005 Rev AA.0