2. Plug the interface cable into the applicable connection (patient station, head wall, or docker station).
3. Press the N N u u r r s s e e c c a a l l l l button (B) to verify the connection between the product's nurse call signal and the hospital's nurse
call system ( Operator control panel, basic, outside siderail (page 26), Operator control panel, advanced, outside siderail
(option) (page 27)).
To activate nurse call communication, see Activating nurse call (page 25).
S S e e t t t t i i n n g g u u p p i i B B e e d d W W i i r r e e l l e e s s s s ( ( o o p p t t i i o o n n ) )
W W A A R R N N I I N N G G - - Always map the i i B B e e d d L L o o c c a a t t o o r r or S S e e c c u u r r e e C C o o n n n n e e c c t t to the location to provide location information. If you move
an i i B B e e d d L L o o c c a a t t o o r r or S S e e c c u u r r e e C C o o n n n n e e c c t t after it has been set up and mapped, you must remap to the new location.
To setup your product to receive a wireless connection, you must set up the i i B B e e d d L L o o c c a a t t o o r r ™ or S S e e c c u u r r e e C C o o n n n n e e c c t t on the
wall at the head end of the product. The i i B B e e d d L L o o c c a a t t o o r r or S S e e c c u u r r e e C C o o n n n n e e c c t t communicates with the product. For instructions
about how to mount the i i B B e e d d L L o o c c a a t t o o r r or S S e e c c u u r r e e C C o o n n n n e e c c t t , see the i i B B e e d d L L o o c c a a t t o o r r Installation Manual or S S e e c c u u r r e e C C o o n n n n e e c c t t
Installation/Configuration Manual.
Contact Stryker Technical Support at (800) 327-0770 with any set up questions.
N N o o t t e e - - You must load the wireless connection settings before the device will communicate with the i i B B e e d d Server application.
See the i i B B e e d d Server Installation/Configuration Manual.
S S e e t t t t i i n n g g u u p p c c a a b b l l e e - - f f r r e e e e n n u u r r s s e e c c a a l l l l c c o o m m m m u u n n i i c c a a t t i i o o n n ( ( o o p p t t i i o o n n ) )
When paired with S S e e c c u u r r e e C C o o n n n n e e c c t t the product can send a cable-free nurse call signal. This setup will be performed by
authorized personnel before the product is placed into service.
To test wireless nurse call communication:
1. Place the product in a position of -6° reverse Trendelenburg or less.
2. Apply the brakes ( Applying or releasing the brakes (page 19)).
N N o o t t e e
• Apply the brakes for S S e e c c u u r r e e C C o o n n n n e e c c t t and the product to establish a first connection.
• Make sure that the product is in a position of -6° reverse Trendelenburg or less to establish a first connection.
• If the product is equipped with the patient helper option, the product must be in a flat position and either be against
the wall at a 14 in. (35.6 cm) product height (or more) or within 3 ft (0.9 m) from the wall at a 22 in. (55.9 cm) product
height (or more) to establish a first connection.
3. Press the N N u u r r s s e e c c a a l l l l button (B) to verify the connection between the product's nurse call signal and the hospital's nurse
call system ( Operator control panel, basic, outside siderail (page 26), Operator control panel, advanced, outside siderail
(option) (page 27)).
To activate nurse call communication, see Activating nurse call (page 25).
EN
16
F F i i g g u u r r e e 4 4 – – 3 3 7 7 - - p p i i n n c c o o n n n n e e c c t t o o r r
3009-009-005 Rev AA.0