N N o o t t e e - - Nurse call requires a connection between the product and an applicable input (patient station, head wall, or docker
station). See Setting up wired nurse call communication (page 15).
C C o o n n n n e e c c t t i i n n g g p p e e r r i i p p h h e e r r a a l l e e q q u u i i p p m m e e n n t t t t o o t t h h e e a a u u x x i i l l i i a a r r y y o o u u t t l l e e t t
W W A A R R N N I I N N G G
• Only use hospital grade electric equipment consuming 5A or less with the auxiliary outlet. The use of standard electric
equipment may bring the current leakage to a level unacceptable for hospital equipment.
• Do not use the auxiliary outlet for life sustaining equipment.
The auxiliary outlet is a built-in outlet for peripheral equipment. The auxiliary outlet (J) is at the foot end of the product
( Product illustration (page 12)).
N N o o t t e e - - Resettable circuit breakers at the head end of the product protect the auxiliary outlet.
C C o o n n n n e e c c t t i i n n g g a a p p a a t t i i e e n n t t d d e e v v i i c c e e t t o o t t h h e e U U S S B B p p o o r r t t ( ( o o p p t t i i o o n n ) )
N N o o t t e e - - The USB port is designed to accommodate devices that comply with the appropriate standards (i.e. IEC 60601-1,
IEC 60950) and rated for a maximum of 5Vdc.
The USB port is a built-in port for patient devices. The USB port is located in the patient device holder (C) on the patient left
siderail ( Product illustration (page 12)).
O O p p e e r r a a t t o o r r c c o o n n t t r r o o l l p p a a n n e e l l , , b b a a s s i i c c , , o o u u t t s s i i d d e e s s i i d d e e r r a a i i l l
W W A A R R N N I I N N G G
• Always lower the product to its lowest height when the patient is unattended to reduce the risk of injury due to patient
falls.
• Always lock the controls when the patient is unattended.
• Always route the cables, wires, and tubing from other equipment so that parts of the product do not pinch them.
C C A A U U T T I I O O N N - - Always clear the product of obstacles before you use motion functions.
N N o o t t e e - - The motion button flashes when the product reaches the motion limit.
A
B
EN
26
C
E
H
D F
G
J
I