Scale system accuracy (NAWI)
MAX = 250 kg, MIN = 20 kg, e = 2 kg, Tare = -60 kg
Patient sleep surface
Bed height to top of seat litter
Gatch position
Fowler position
Trendelenburg and reverse Trendelenburg
Electrical requirements
N N o o t t e e - - Class I Electrical Equipment: Protection against electrical
shock relies on connection to protective earth of an appropriately
rated hospital grade outlet.
Hospital grade auxiliary outlet
Battery voltage
N N o o t t e e - - Always replace with Stryker approved batteries.
Duty cycle
Application environments
Maximum acoustic sound pressure
C C o o m m p p a a t t i i b b l l e e
L L e e n n g g t t h h
s s u u p p p p o o r r t t
s s u u r r f f a a c c e e s s
Model 2850
84 in.
C C o o m m f f o o r r t t G G e e l l ™
Model 2860
84 in.
I I s s o o F F l l e e x x ®
Model 2872
84.25 in.
I I s s o o T T o o u u r r ™
Model 2815
84 in.
P P r r o o F F o o r r m m ®
Model 2940/
84 in.
2941 I I s s o o A A i i r r ®
Stryker reserves the right to change specifications without notice.
Specifications listed are approximate and may vary slightly from product to product or by power supply fluctuations.
EN
8
Standard
Bed extender
Standard
Zoom
W W i i d d t t h h
213.4 cm
35 in.
213.4 cm
35 in.
214 cm
35.5 in.
213.4 cm
35 in.
213.4 cm
35 in.
± 3% of the total patient weight for patients who
weigh 100 lb (45.4 kg) to 550 lb (249.5 kg)
± 2.2 lb (1 kg) for patients who weigh 44 lb (20
kg) to 220 lb (100 kg)
± 4.4 lb (2 kg) for patients who weigh 220 lb (100
kg) to 551 lb (250 kg)
84 in. x 35 in.
96 in. x 35 in.
11.5 in. to 30 in.
14 in. to 32 in.
0° to 30° ± 5°
0° to 65° ± 5°
+12° to -10° ± 5°
120 VAC, 60 Hz, 8A
120 VAC, 60 Hz, 8A
12 VDC, 1.2 Ah (x2) (Stryker part number:
700000341245)
2 minutes ON, 18 minutes OFF
1, 2, 3, and 5 per IEC 60601-2-52
64 dBa
T T h h i i c c k k n n e e s s s s
88.9 cm
7 in.
88.9 cm
6 in.
90.2 cm
9.5 in.
88.9 cm
6 in.
88.9 cm
7 in.
213.4 cm x 88.9 cm
243.8 cm x 88.9 cm
29.2 cm x 76.2 cm
35.6 cm x 81.3 cm
230 VAC, 50 Hz, 8A
230 VAC, 50 Hz, 8A
17.8 cm
15.2 cm
24.1 cm
15.2 cm
17.8 cm
3009-009-005 Rev AA.0