Malfunctions and troubleshooting

Malfunction
Control unit is
not switched on
Power connection not
established
Incorrect operating voltage Check the mains voltage
Faulty fuse
Motor not switched on
Incorrect motor active
Motor not connected
Handpiece or contra angle
not correctly assembled
Peristaltic pump not
switched on
Tubing set incorrectly
inserted
Tubing set clogged/crusted
matter visible
Bottle with sodium chloride
solution not ventilated
Tubing set is dripping
Roller clamp of tubing
set is closed
Pedal not connected
Incorrect operation
manuals search engine
Cause
Solution
Set the
power switch "I/O"
to "I"
Connect the control unit to the
mains power supply
Replace the fuse
Switch on the motor using the
treadplate
Switching to the other motor,
using the Vario pedal
Connect the motor cable to the
control unit
Press the handpiece firmly onto
the electronic motor until it
clicks into position and check
that it is secure by moving it
slightly in the opposite direction.
Switch on the peristaltic pump
Insert tubing set correctly (note
the direction)
Replace the tubing set
Open the ventilation filter at the
drip chamber
Replace the tubing set
Open roller clamp all the way
Connect the pedal to the control
unit
Check operating instructions
Refer to operating
instructions
7.1 Switching the device on and
off
6.2 Connection to the power
supply
6.2 Connection to the power
supply
9.1 Replacing the control unit fuse
7.8 Operation using the Vario
pedal
7.8 Operation using the Vario
pedal
5.0 Device overview
6.2 Connection to the power
supply
6.3 Device preparation
7.8 Operation using the Vario
pedal
6.3 Device preparation
6.3 Device preparation
6.3 Device preparation
6.3 Device preparation
6.3 Device preparation
6.3 Device preparation
7.8 Operation using the Vario
pedal
27
EN

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis