Parameter Level 2
Values in Parameter Level 2 can only be adjusted after the password "9403" was entered. The password cannot be changed.
Entering password:
Handpiece
Max. Speed
Handpiece max speed/HP 01
Handpiece max speed/HP 02
Handpiece max speed/HP 03
Handpiece max speed/HP 04
Handpiece max speed/HP 07
Handpiece max speed/HP 08
Handpiece max speed/HP 09
Handpiece max speed/HP 10
Handpiece max speed/HP 11
Handpiece max speed/HP 12
Handpiece max speed/HP 13
Handpiece max speed/HP 14
Handpiece max speed/HP 15
* Speed selection in 4 steps: 9000, 10'000, 11'000, 12'000 rpm
Pump parameters
Pump/Backwards turn mode variable
Pump/Way backwards
Pump/Speed backwards
Pump/Range 1 increment
Pump/Range 1 end
Pump/Range 2 increment
Pump/Range 2 end
Pump/Range 3 increment
Pump/Range 3 end
Pump switch mode on/off
HighSurg 30 Operating Instructions, Ref. 31666, V24/16
Downloaded from
www.Manualslib.com
Press "Enter", press "Select
To confirm press "Enter" for 1 second, until a tone is emitted.
Name of handpiece
on the display
1 : 5
1 : 3
1 : 2
1 : 1
Kirschner Hp
Tattoo Hp
Cont. Shaver
Osc.Shaver
Jacobs Chuck
Perforator
Craniotome
4 : 1
OTO-Drill Hp
manuals search engine
or
", (for fast forward or backward keep key pressed).
Speed range
Factory
rpm
default
1500 – 240,000 240,000
900 – 150,000
150,000
600 – 100,000
100,000
300 – 80,000
80,000
500 – 2800
2800
9000 – 12,000* 12,000
300 – 6000
6000
300 – 5000
5000
200 – 2600
2600
80 – 1200
1200
1000 – 60,000
60,000
75 – 20,000
20,000
300 – 16,000
16,000
Range
Factory
Definition
default
No/Yes
Yes
The pressures in the tube set vary according to
the pump speed. In "variable mode" this is
considered, to prevent of spilling when pump
is switched off in backwards mode.
1 – 100%
25%
Specify how far the pump turns backwards
10 – 50%
33%
Specify how fast the pump has to turn
backwards to prevent of spilling after the
switch off of the motor.
1 – 10%
1%
Adjustment steps in section 1.
5 – 50%
20%
Set the range where section 1 is active
1 – 10%
5%
Adjustment steps in section 2
10 – 90%
50%
Set the range where section 2 is active.
1 – 10%
10%
Adjustment steps in section 3
20 – 100%
100%
Set the range where section 3 is active.
No/Yes
No
No: Pump starts simultaneously with the
Yes: Pump starts separately by pressing the
Definition
Limit the maximum speed of your
handpieces according to your own
experience.
handpiece drive.
foot switch.