Herunterladen Inhalt Inhalt Diese Seite drucken

Operation Of Conform Cannula Handpiece; Emptying Secretion Jar; Function Control; General Functions - Nouvag Vacuson 60 LP Bedienungsanleitung

Liposuction system
Vorschau ausblenden Andere Handbücher für Vacuson 60 LP:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

7.5

Operation of Conform Cannula Handpiece

By pressing the foot switch "Motor", the stroke movement of the Conform Cannula Handpiece is actuat-
ed. The motor is running with a constant speed of 10,000 rpm.
7.6

Emptying secretion jar

The jar lid of the secretion jar is equipped with an overflow protection system to prevent the vacuum sys-
tem from being flooded by secretion fluids at high filling levels of the secretion jar. Therefor a float gauge is
responsible.
At high filling levels of the secretion jar and the resulting locking of the overflow protection system the se-
cretion jar has to be emptied or replaced by another secretion jar.
1. Swich off suction pump.
2. Disconnect tubes including connectors from the secretion jar lid.
3. Unhinge full secretion jar from the secretion jar mount and dispose of secretion fluids according to na-
tional disposal regulations.
4. Procure used secretion jars to the reprocessing cycle.
5. Connect suspended tubes with new, ready to use secretion jar.
7.7

Function control

To obtain trouble-free operation of the suction pump all the components and functions of the pump system
have to be tested prior every assignment.

General functions:

1. Use the power switch "I/O", at the rear of the device, to switch the device on, indicator light, LED, is il-
luminated.
2. Neither the Infiltration pump nor the suction pump or the motor of the Conform Cannula Handpiece
may be actuated without the pedal being pressed.
Suction pump:
1. Press foot switch "Vacuum" (blue) at the VARIO pedal to activiate the suction pump.
2. Hold hand in front of the "Exhaust" (Rear of the device). Airflow is perceptible.
3. Turn Vacuum Controller (Vacuum) clockwise, equivalent to vacuum maximum.
4. Control vacuum performance at the openings of the cannula. Suction performance is high.
5. Turn Vacuum Controller anticlockwise all the way to the stop, equivalent to vacuum minimum.
6. Control vacuum intensity at the openings of the cannula. Suction performance is low.
7. Use the power switch "I/O", at the rear of the device, to switch off device, indicator light, LED, goes off.
Infiltration pump:
1. Turn Infiltration Controller, "Infiltration", clockwise, all the way it goes, equivalent to maximal flow rate.
2. Press pedal on the VARIO Pedal, infiltration pump starts.
3. Turn Infiltration Controller couterclockwise, all the way it goes, equivalent to minimum flow rate.
4. Infiltration performance is reduced down to a stand still.
5. Once the maximum infiltration amount is set, the delivery of infiltration liquid can be regulated by pe-
dal.
Conform Cannula Handpiece with Electronicmotor:
By pressing the foot switch "Motor" (yellow) the electronic motor of the Conform Cannula Handpiece is ac-
tivated and runs with a constant speed. Vibrations at the Conform Cannula Handpiece are perceptible.

Malfunctions and troubleshooting:

To solve problems refer to chapter 10 "Malfunctions and troubleshooting".
16
Operating instructions Vacuson 60 LP, REF 31793, V34/17
Vacuson 60 LP

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis