  • DEUTSCH, seite 1
HighSurg 30, Error-messages on Display
Error-messages/Error Code
Basic Initialization/
Woo
Set default value/
Wo1
Memory error/
Eo2
Handling error/
E03
Program SW error/
E04
UserConfig SW error/
E05
Display error/
E06
Pump error/
E07
Storing factory settings/
User Config & Program
Storing factory settings/
Program
Pedal not connected/
E10
Pedal test mode/
W11
Keyboard test mode/
W12
No motor connected/
E13
Motor 2 not connected/
E14
Motor 1 not connected/
E15
Unknown motor 2/
E16
Unknown motor 1/
E17
28
Downloaded from
www.Manualslib.com
Cause
First Initialization
Parameter reset to default value
System Error
System Error
System Error
System Error
System Error
System Error
Message while default values of parameters
and programs are saved to the NOU-dongle.
Message while default values of programs are
stored.
a) Pedal is not plugged in.
b) Plug or cable is defective.
Pedal test mode switched on
Keyboard test mode switched on
a) No motor connected
b) Motor, motor cable, motor plug or Control
Unit are is defective
a) Motor 2 is selected but no motor plugged
in
b) Motor connected to motor socket 2, but
motor, motor cable, motor plug or Control
Unit is defective.
a) Motor 1 is selected but no motor plugged
in.
b) Motor connected to motor socket 1, but
motor, motor cable, motor plug or Control
Unit is defective.
a) Motor 2 is selected but wrong motor is
plugged in.
b) Right motor is connected to motor socket
2, but motor, motor cable, motor plug or
Control Unit is defective.
a) Motor 2 is selected but wrong motor is
plugged in.
manuals search engine
Solution
Send Control Unit to Service Center.
Send Control Unit to Service Center.
Send Control Unit to Service Center.
Send Control Unit to Service Center.
Send Control Unit to Service Center.
Send Control Unit to Service Center.
a) Plug in Pedal
b) Send Control Unit and pedal to Service
Center.
Switch off device for 5 seconds, than switch
on again.
Switch off device for 5 seconds, than switch
on again.
a) Plug in motor
b) Send motor and Control Unit to Service
Center.
a) Plug in motor
b) Send motor and Control Unit to Service
Center.
a) Plug in motor
b) Send motor and Control Unit to Service
Center.
a) Plug in correct motor
b) Send motor and Control Unit to Service
Center.
a) Plug in correct motor.
b) Send motor and Control Unit to Service
HighSurg 30 Operating Instructions, Ref. 31666, V24/16