Cleaning, disinfection and sterilization
The instructions described here are intended for the parts supplied in the set. The cleaning, disinfection and
sterilization instructions for extensions and accessories are described in their respective operating
instructions.
The following points in particular are important with regard to caring for the material:
● Perform cleaning, disinfection and sterilization after every treatment.
● Always autoclave the material in sterilization packaging.
● Make sure that sterilization packaging is no more than 80 % full.
● Always autoclave the material at 135°C for at least 5 minutes.
● If sterilized material is not used immediately, the material packaging must be labeled with the
sterilization date.
● Nouvag AG recommends including a sterility indicator.
Control unit and pedal
The control unit and pedal do not come into contact with the patient.
Wipe the outside using micro-biologically tested surface disinfectant or a 70 % isopropyl alcohol solution.
The front plate of the control unit is sealed for this purpose and can be wiped clean.
OTO-Drill handpiece with integrated Electronic Motor 16
The OTO-Drill handpiece is can be ordered optionally. The reprocessing instructions are contained in the
operation manual, delivered with the product.