7.2
A) Display: Shows the operating values (See chapter "7.3 Overview: Standard display").
B) "Program" key: Selection of program 1 to 10. There are multiple programs for each motor selectable.
(To facilitate the selection of programs, the selectable programs can be limited to the number you need. Refer to
"7.7 Configuration Menu, Level 1, Programs/Number of programs").
C) "Motor 1/2" key: Switching between the connected motors (the green indicator lights beside the motor sockets
indicate the active motor). Pressing the Motor key longer changes direction of rotation.
D) "Handpiece" key: Choice of handpieces or contra angle (can be individually deactivated, refer to
"7.7 Configuration Menu").
E) "Enter" key: Used in the configuration menu (refer to "7.7 Configuration Menu")
F) "Select" keys:
G) "Speed" keys:
H) "Torque/Frequency" keys:
I)
"Pump" keys:
HighSurg 30 Operating Instructions, Ref. 31666, V24/16
Downloaded from
www.Manualslib.com

Operation

Switching the device on and off

The power switch "I/O" (at the rear) is used to switch the control unit on and off.
The device can be switched off at any time irrespective of any procedure for switching off the device.

Overview: Control elements on the operation panel

A
By pressing the left "Select" key "Select ▲" the software version is shown.
By pressing the right key "Select ▼", the type of connected motor is displayed.
By pressing both "Select" keys at the same time "Select ▲ + ▼" the programs will be reset to factory default
settings.
The Select keys serve in the configuration menu to change values and parameters
"▲": value adjustment (upwards)
Restrict the maximum speed that can be selected using the pedal.
"▲": increases the maximum speed
By pressing both "Speed" keys at the same time "Speed ▲ + ▼" the handpiece calibration will be started
(refer to "7.4.2 Calibrating handpieces").
Restricts the maximum torque.
"▲": increases the maximum torque
With the shaver handpiece attached the key pair is used to set the oscillation period, if oscillation mode is
selected.
Changing the pump flow rate that can be supplied using the pedal.
"▲": increases the maximum supply quantity
By pressing both "Pump" keys at the same time "Pump ▲ + ▼" the pump will be put on call, pressing again
will switch it off.
manuals search engine
B
C
D
E
"▼": value adjustment (downwards)
"▼": reduces the maximum speed
"▼": reduces the maximum torque
"▼": reduces the maximum supply quantity
F