Motor system for medical surgery
Vorschau ausblenden Andere Handbücher für HighSurg 30:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
HighSurg 30
Hp not allowed for motor 2/
E18
Hp not allowed for motor 1/
E19
Pump is open/
E20
Motor or pump test mode/
W21
Pedal locked/
W26, pedal let go
Battery is almost empty/
W27, continue with „Enter"
Watch Error XX/
E28, continue with „Enter"
Handpiece XX is faulty/
E29
Handpiece XX is Ok!
Testing the handpiece XX
Check max. speed
handpiece/
W30, wait 1 second
NOU-Dongle is plugged in
The red highlighted error messages are also illuminated red on the control device display. The other messages are for
information and do not require any action by the user.
HighSurg 30 Operating Instructions, Ref. 31666, V24/16
Downloaded from
www.Manualslib.com
b) Right motor is connected to motor socket
1, but motor, motor cable, motor plug or
Control Unit is defective.
The Handpiece may not be operated with the
motor plugged in at the motor socket 2.
The Handpiece may not be operated with the
motor plugged in at the motor socket 1.
Motor is not working when pump
compartment is open to prevent of injuries.
Motor or pump test mode is switched on.
If pedal is pressed at switch on procedure,
pedal will not work.
Battery is almost empty.
a) Clock at Control Unit is defective.
b) Device was switched on after battery
change but clock is not set yet.
At calibrating or testing handpiece was
overexposed to high torque.
Tested handpiece is OK.
Handpiece is testing.
Message is displayed when motor with
80,000 rpm is selected and max. speed is set
to 50,000 rpm. Check Handpiece for max
speed clearance before you proceed.
Message is displayed for 1 second after
NOU-dongle was plugged in.
manuals search engine
Center.
Choose appropriate Handpiece or change to
correct motor type.
Choose appropriate Handpiece or change to
correct motor type.
Close pump compartment.
Switch off device for 5 seconds, than switch
on again.
Release pedal for one second.
-
After pressing „Enter", work can be
continued.
-
Send Control Unit as soon as possible to
Service Center.
a) After pressing „Enter", work can be
continued but Control Unit has to be
sent to Service Center as soon as
possible.
b) Set date and time.
-
Clean handpiece and spray it thoroughly
with Nou-Clean spray.
-
If message is still displayed after test
procedure, handpiece has to be sent to
Service Center.
Check handpiece and wait for the message
to disappear. Motor starts if the pedal still is
pressed.
EN