Herunterladen Inhalt Inhalt Diese Seite drucken

Control Unit And Foot Control; Micromotor 31S; Troubleshooting; Disposal - Nouvag TCM 3000 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

NOUVAG AG

6.1 Control unit and foot control

The control unit and the foot control do not come in contact with patients. Clean the surface of the unit
only, do not use harsh cleaners or solvents for cleaning. Use 80% ethyl alcohol or microbiologically
effective disinfectants. The control unit face is sealed and washable.
The control unit should be wiped with a clean, damp cloth only. No spray cleaners
or disinfecting agents should be employed, as the control unit face is not
watertight!

6.2 Micromotor 31S

1)
The micromotor, cable and plug should be cleaned after each use to prevent the build up of
deposits and debris that can destroy it, if not removed. Wipe it with a clean cloth dampened with
disinfectant solution.
2)
Pack micromotor in transparent bag.
3)
After packing motor and cable, autoclave it at maximum 134° C .
4)
After autoclaving, allow electronicmotor to dry for one hour at room temperature.
Do not clean micromotor with compressed air!
After autoclaving allow micromotor to dry for one hour at room temperature.
Do not bend the micromotor cable!
Problem
Motor does not run
Foot control fails to function
Micromotor does not run
In the event a problem cannot be solved, contact dealer or authorized service technicians, as listed on
the last page of your Operating Instruction Manual.
Disposal of device, components and accessories must strictly conform to local laws and regulations as
set out by the relevant authorities.
With regard to the preservation of the environment old equipment may be returned to the distributor or
manufacturer.
No. 31891 31/05

7 Troubleshooting

Cause:
Plug is not inserted properly
Wrong operating voltage
Foot control not connected
Control unit not turned on
Motor not turned on
Motor is not connected
Motor is not connected properly

8 Disposal

TCM 3000
Solution:
Insert plug and check fitting
Check mains voltage output → see
device label
Plug foot control into control unit
Turn main switch "ON/OFF"ON
Turn "motor" with button or with pedal
on
Connect motor cable to control unit
Press micromotor firm on
handpiece until handpiece snaps in
7

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis