Herunterladen Inhalt Inhalt Diese Seite drucken

Parameter Level - Nouvag HighSurg 30 Bedienungsanleitung

Motor system for medical surgery
Vorschau ausblenden Andere Handbücher für HighSurg 30:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1

Parameter Level 1

Group/Parameter
Software/Version
Hardware/Serial number MB
Date-Time/Date format US
Date-Time/Date
Date-Time/Time
Battery/Voltage
Language/0 = DE, 1 = EN
Backlight/brightness (0. .10)
Program/Number of programs
for 50,000 rpm motor
Program/Number of programs
for 80,000 rpm motor
Program/Number of programs
for 16,000 rpm motor
Program/Power on at last
program
Operating hours/Control Unit
Operating hours/Motor 1
Operating hours/Motor 2
Operating hours/Pump
Error memory/ 1 – 8
Handpiece activation
Handpiece existing/HP 01
Handpiece existing/HP 02
Handpiece existing/HP 03
Handpiece existing/HP 05
Handpiece existing/HP 06
Handpiece existing/HP 07
Handpiece existing/HP 08
Handpiece existing/HP 09
Handpiece existing/HP 10
Handpiece existing/HP 11
Handpiece existing/HP 12
Handpiece existing/HP 13
Handpiece existing/HP 14
Handpiece existing/HP 15
18
Downloaded from
www.Manualslib.com
Rights
read
read
read/write
read/write
read/write
read
read/write
read/write
read/write
read/write
read/write
Read/write
read
read
read
read
read
Name of handpiece
on display
1 : 5
1 : 3
1 : 2
Com.Saw
Osc.Saw
Sag.Saw
Dermatome
Kirschner Hp
Tattoo Hp
Cont.Shaver
Osc.Shaver
Jacobs Chuck
Perforator
Craniotome
manuals search engine
Factory
Definition
VX.XX
Shows current software version
92XXXXXXX
Shows serial number of the main board
no
Set to US Date format
Change current date
Change current time
e.g. 3120 mV
State of charge of battery
(Capacity about 3100 mV, Low 1800 mV)
1
Change display language between German and English
9
Display brightness, variable: 0, ... , 10
10
Number of activated Programs: 3, ... , 10
6
Number of activated Programs: 1, ... , 7
1
Number of activated Programs: 1, ... , 3
Yes
No: Display of Motor 1 with Program 1
Yes: Display of the last used program/motor
XX:XX:XX
Shows operating hours of HighSurg 30
XX:XX:XX
Shows operating hours of motor 1
XX:XX:XX
Shows operating hours of motor 2
XX:XX:XX
Shows operating hours of pump
X
8 Error messages in chronological order.
For the list of error codes please refer to chapter 10.
Choice
Factory
default
yes/no
yes
yes/no
yes
yes/no
yes
yes/no
yes
yes/no
yes
yes/no
yes
yes/no
yes
yes/no
yes
yes/no
yes
yes/no
yes
yes/no
yes
yes/no
no
yes/no
no
Yes/no
no
HighSurg 30 Operating Instructions, Ref. 31666, V24/16
HighSurg 30
Definition
Deselect handpieces that do not belong
to your product range by switching to
"no". This will later shorten the scrolling
list for "Handpieces". Otherwise you will
later have to scroll down the complete
list of available handpieces, each time
you select a handpiece.
Because HP 04, 1:1 is permanently active,
it's not listed in this table.

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis