Herunterladen Inhalt Inhalt Diese Seite drucken

Nouvag HighSurg 30 Bedienungsanleitung Seite 43

Motor system for medical surgery
Vorschau ausblenden Andere Handbücher für HighSurg 30:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
HighSurg 30
HighSurg 30 Operating Instructions, Ref. 31666, V24/16
Downloaded from
www.Manualslib.com
A
A) Press release key for tubing set bracket (on top of the control unit) to open the pump.
B) The compartment with the integrated tubing bracket opens.
C) Place the tubing set into the tubing bracket provided in such a way that the part of the tubing set
with the spike exits the pump towards the rear of the device. Make sure the tubing is secure.
D) With the tubing set inserted, press the compartment downwards until it clicks into place.
8
8.
Insert the spike at the end of the tubing set into the irrigation fluid bottle and hang the bottle onto the
stand.
9.
Open the roller clamp on the tubing set as far as it will go.
10. Open the bleed valve beneath the drip chamber.
11. Connect the control unit to the power socket.
Ensure that the operating voltage setting corresponds to the local mains voltage.
manuals search engine
B
9
D
C
10
EN
9

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis