Number of available storing places for each motor and both motor connections
Motor-Choice
Motor 50,000 rpm
Motor 80,000 rpm
Motor 16,000 rpm, OTO-Drill
16
Downloaded from
www.Manualslib.com

Step 5: Setting the pump supply quantity

Press the "Pump" keys "▲" to increase or "▼" to decrease the pump supply quantity. By pressing the key
continuously the values are fast forwarded.
Minimum and maximum of pump supply quantity as well as operating steps can be adjusted in the
Configuration Menu (refer to «7.7 Configuration Menu, Parameter level 2, Pump»).
To activate or deactivate the pump, press both "Pump" keys at the same time,
"Pump ▲ + ▼", or use the foot switch

Torque limiter

The device has an automatic torque limiter, which functions like a torque wrench.
If the attached instrument meets resistance, the torque increases as far as the maximum level, and the
speed then reduces, to standstill if necessary. The torque on the instrument is maintained. If the load on the
instrument reduces, the speed increases again up to the set maximum level.
This process can easily be seen on the display by means of the bar chart. The segments of the bar chart fill
up as the resistance on the instrument increases. When the torque has reached the set maximum, that is,
when all segments are visible, the speed reduces. As soon as the pressure on the instrument is reduced, the
torque falls again. The bar chart on the display regresses and the instrument speed increases again.

Programme memory

At switch off of the HighSurg 30 all the settings made by the user are automatically stored in each program
place. The display shows every setting of the selected program place. The following parameters are
automatically stored:
• Handpiece/Transmission ratio
• Maximal speed
• Maximal torque
To change a program go to the specific parameter and change it. When the device is switched off all
parameters are automatically stored under the same program number.
In the configuration menu the number of storable programs can be limited for each motor separately.
In the configuration menu it can be chosen what the display shows at start-up of the device. Choose
between Program 1 with Motor 1 or the last used setting before the device was switched off.
Number of available
storing places
3 – 10
1 – 7
1 – 3
manuals search engine
.
• Pump On/Off
• Pump performance
• Shaver oscillation period
Motor connection
Factory
default setting
10
6
1
HighSurg 30 Operating Instructions, Ref. 31666, V24/16
HighSurg 30
Motor connection
Number of available
Factory
storing places
default setting
3 – 10
10
1 – 7
1 – 3
6
1