The HighSurg 30 can recognize the type of a plugged-in motor. This enables the adaption of the motors and their
safe operation.
Motor 21, 50,000 rpm
Motor 21, 50,000 rpm/Min. speed
Motor 21, 50,000 rpm /Max. speed
Motor 21, 50,000 rpm/Ramp start
Motor 21, 50,000 rpm/Ramp stop
Motor 21, 80,000 rpm
Motor 21, 80,000 rpm/Min. speed
Motor 21, 80,000 rpm /Max. speed
Motor 21, 80,000 rpm/Ramp start
Motor 21, 80,000 rpm/Ramp stop
Max. speed warning delay
Motor 16, OTO-Drill
Motor 16, OTO-Drill/
Min. speed
Motor 16, OTO-Drill/
Max. speed
Motor 16, OTO-Drill/
Ramp start
Motor 16, OTO-Drill/
Ramp stop
Resetting to
factory default
Default value/Set default value
20
Downloaded from
www.Manualslib.com
Range
300 – 5000 rpm
5000 – 50,000 rpm
1 – 1000 ms/10,000 rpm 250 ms
1 – 1000 ms/10,000 rpm 50 ms
Range
300 – 5000 rpm
5000 – 80,000 rpm
1 – 1000 ms/10,000 rpm 250 ms
1 – 1000 ms/10,000 rpm 50 ms
0 – 1000 ms
Range
1000 – 5000 rpm
5000 – 16,000 rpm
1 – 1000 ms/10,000 rpm 250 ms
1 – 1000 ms/10,000 rpm 50 ms
Attention: Upon resetting to factory default all parameters will appear with factory default values
(except date and time and the operating hours counter).
Note: For resetting the "programs" to factory default press both "Select" keys at the same time
("Select ▲ + ▼"). For this procedure you have to be out of the Configuration Menu.
manuals search engine
Factory
Definition
default
300 rpm
Set the min. speed
50,000 rpm Set max. speed of
Set acceleration time to 10,000 rpm
Set breaking time from 10,000 – 0 rpm
Factory
Definition
default
300 rpm
Set the min. speed of
80,000 rpm Set max. speed of
Set acceleration time to 10,000 rpm
Set breaking time from 10,000 – 0 rpm
1000 ms
0:
1 – 1000: Delay time of warning when
Factory
Definition
default
1000 rpm
Setting of the min. speed
16,000 rpm
Setting of max. speed
Set acceleration time to 10,000 rpm
Set breaking time from 10,000 – 0 rpm
Choice
Factory
Definition
default
Yes/No
Resetting all parameters of the Configuration
Menu to factory default.
HighSurg 30 Operating Instructions, Ref. 31666, V24/16
HighSurg 30
Warning is switched off
exceeding 50,000 rpm.