Malfunction and troubleshooting

Malfunction
Mesh board is not feeding
Split skin transplant can
not, or not all the way, be
stretched all the way.
Mesh board is worked
thru the Mesher but the
split skin transplant is not
on it.
If a fault cannot be rectified, please contact your supplier or an authorized service center. The addresses are provided on
the last page of this operation manual.
11

Consumables and spare parts

Parts
Skin transplant mesh board with a mesh factor of 1.5 in PU of 20 pieces ------------------------------------------------------- 1981
Skin transplant mesh board with a mesh factor of 3.0 in PU of 20 pieces ------------------------------------------------------ 1982
Skin transplant mesh board with a mesh factor of 6.0 in PU of 20 pieces ----------------------------------------------------- 2105
Skin transplant mesh board with a mesh factor of 9.0 in PU of 20 pieces ----------------------------------------------------- 2106
Spare blade roll ------------------------------------------------------------------------------------------------------------------------------- 75591
Non slip silicone mat, 450 x 270 mm ---------------------------------------------------------------------------------------------------- 4126
Spare paper filter for Stericase in PU of 100 pieces ---------------------------------------------------------------------------------- 4127
Spare ratchet with case ------------------------------------------------------------------------------------------------------------------ 30444
Mesher operation instructions on CD-ROM ---------------------------------------------------------------------------------------- 31584*
*At Nouvag AG the operation instructions are delivered as a PDF-File on a CD-ROM, together with the according device. If
you prefer a printed version of it, we will send you one, free of charge, by postal delivery. If you cannot find the manual
anymore, we'll send you a spare one in PDF-Format by E-Mail.
To order additional parts, please contact our customer service department.
Operating instructions Mesher, Ref. 31584, V32/15
Cause
Roller pair did not catch the
mesh board.
Backing roll builds up too
little backpressure
Backing roll builds up too
little backpressure
Split skin transplant got stuck
on the blade roll.
Solution
Press mesh board with a
slight pressure against the
roller pair.
Adjust pull-twist-knob to a
smaller roller distance.
Adjust pull-twist-knob to a
smaller roller distance.
After the first pressing of the
ratchet check whether the
split skin is stuck on the blade
roll. Lay the split skin back on
the mesh board
Reference in the
operation manual
7.2 Use of the Mesher
6.2 Device preparation
6.2 Device preparation
7.2 Use of the Mesher