6.2 Charging the prosthesis battery

INFORMATION
When the Protective Cover is installed, the battery charger cable has to point to the upper clos­
ure. A correct knee joint charging process is only ensured with this alignment.

6.3 Display of the current charge level

6.3.1 Display of battery charge level without additional devices

INFORMATION
The charge level cannot be displayed during the charging process, e.g. by turning the prosthes­
is over. The product is in charging mode.
Beep signal
5x short
4x short
3x short
2x short
1x short
INFORMATION
If the Volume parameter is set to '0' in the Cockpit app, there are no beep signals (see
Page 84).
74
1) Connect the inductive charger to the receiver of the charging unit
on the rear of the product. The charger is held in place by a mag­
net.
→ The LED ring on the rear of the charger pulsates purple
(4-second cycle).
→ If the LED ring lights up in a different colour, this indicates an
error (see Page 100).
2) The charging process starts.
→ Once the product battery is fully charged, all LEDs on the side
of the battery charger light up.
3) After the charging process is complete, hold the prosthesis still and
remove the inductive charger from the receiver.
→ A self-test is performed. The joint is operational only after cor­
responding feedback (see Page 103).
1) Turn the prosthesis by 180° (the sole of the foot must face up).
2) Hold still for 2 seconds and wait for beep signals.
Battery charge level
more than 80%
60% to 80%
40% to 60%
20% to 40%
less than 20%