5.3. Restoring factory settings
To restore the factory settings (see paragraph 3.5) press keys
simultaneously. "rESS" will briefly appear and the handheld will be rebooted.
6. Error messages
NOTE: In presence of strong wireless communication traffic, the connection between handheld and generator may break down.
To restore the connection run the "Restoring factory settings" procedure.
FAULT
C AU SE
E01
X-RAY KEY RELEASED TOO EARLY
E02
SHOOTING SEQUENCE NOT COMPLETED
E03
HANDHELD INTERNAL TEST ERROR
E04
E05
HANDHELD AUTO DIAGNOSIS TEST FAILED
E08
E06
GENERAL HANDHELD ERROR
E07
RF SIGNAL TOO LOW
E09
HANDHELD SERIAL NUMBER INCORRECT OR NOT
INITIALIZED
E10
E12
X-RAY UNIT INTERNAL ERROR
E13
E16
EN
and
OPERATING INSTRUCTIONS
to go to the set up menu. Press keys
S O L U TION
Hold down the key until the image has been captured.
Handheld most likely lost the signal. Try to repeat
exposure. If the problem persists, contact technical
service.
Take out the batteries and then put them back in after
waiting a few seconds. If the problem persists, contact
technical service.
Contact technical service.
Try to repeat exposure. If the problem persists, contact
technical service.
Handheld lost the signal. Try to repeat exposure. If the
problem persists, contact technical service.
Contact technical service.
Contact technical service.
and
15
15