Electrical connection

Danger due to electric voltage
‣ T he entire process of electrical connection must be carried out while the system is
de-energized.
‣ B efore commissioning, compare the supply voltage to the specifications on the nameplate to
ensure that they match.
‣ P rovide a suitable switch or power-circuit breaker in accordance with IEC/EN61010 within
the building installation. This switch must be fitted near to the device (within easy reach)
and must be marked as a circuit breaker.
‣ An overcurrent protection device (rated current of ≤ 10 A) is required for the power cable.
• Note the terminal designation on the side of the device.
• Mixed connection of safety extra low voltage and voltages that are dangerous to touch
to the relay/SSR is permitted.
In the case of a device for use in a hazardous area:
Note the information in the Ex documentation (XA).
5.1

Connection conditions

5.1.1
Required tool
For the terminals:
0.6 x 3.5 mm
5.1.2
Connecting cable requirements
The connecting cables provided by the customer (power supply and output signal) and
the connection lines (FTR525 to FQR57/FDR57) must meet the following requirements:
• Electrical safety in accordance with applicable national regulations.
• Permitted temperature range → 48
• Power supply and signal lines: Normal installation cable
• Connection line: Normal installation line, at least a three-wire cable
• Line cross-sections: 0.2 to 2.5 mm
• T he terminals of the FTR525 are only approved for connecting fixed lines. The operator
must ensure adequate strain relief.