Nivotester FTR525
7.4.2.3
Current output assignment
Application
Assignment outputs
→
2000
This function is used to assign the current output to measuring channel 1
(limit, 1M) or measuring channel 2 (motion, 2M).
Options: "Limit (1M)", "Motion (2M)"
7.4.2.4
Current output mode
Application
Current output
→
2000
This function can be used to switch on output magnify.
4 mA
Standard
Grenzstand 1
4 mA
Function 2320
30 Current output mode
For this function, a small segment of the 0 to 100 % detection range is shown on the current
output of 4 to 20 mA.
Options: "Standard", "Magnify"
7.4.2.5
4mA value of the magnify
Application
Current output
→
2000
This function is used to enter the percentage value for the lower limit (4 mA) while the
magnify is enabled (function 2310).
Options: 0 to 100 %
7.4.2.6
20mA value of the magnify
Application
Current output
→
2000
This function is used to enter the percentage value for the upper limit (20 mA) while the
magnify is enabled (function 2310).
Options: 0 to 100%
The 4mA value (function 2310) must always be lower than the 20mA value; if it is not,
an error message is displayed.
Endress+Hauser
2200
→
2300
20 mA
082
%
Magnify
20 mA
Function 2330
→
2300
→
2300
Current output
→
2230
Current output mode
2310
4mA value
2320
20mA value
2330
Commissioning
65