XA00250P-E

Temperature tables

Connection data

16
Type of protection
ATEX:
II 2 G Ex db ia IIC T6...T4/T3 Gb
IECEx:
Ex db ia IIC T6...T4/T3 Gb
1)
Only in connection with Basic specification, Position 10 + 11 (Additional Option 1 + 2) = T;
see Operating Instructions
• The specified ambient and process temperature ranges exclusively refer to the explosion
protection and must not be exceeded. Operationally permitted ambient temperature ranges
can be restricted depending on the version: See Operating Instructions.
• Do not exceed the max. ambient temperature at the housing.
Basic specification, Position 2 (Output, Operating) = A, B, C, D, E, F
Power supply
U ≤ 45 V
DC
P ≤ 3 W
Basic specification, Position 2 (Output, Operating) = M, N, O, P, Q, R
Power supply
U ≤ 32 V
DC
P ≤ 3 W
Temperature
Process temperature T
class
(process)
T6
–40 °C ≤ T
≤ +75 °C
p
T4
–40 °C ≤ T
≤ +100 °C
p
T4
–40 °C ≤ T
≤ +125 °C
p
T4
–40 °C ≤ T
≤ +125 °C
p
T3
–40 °C ≤ T
≤ +150 °C
p
Cerabar S PMC71
Ambient temperature T
p
(ambient)
–40 °C ≤ T
≤ +40 °C
a
–40 °C ≤ T
≤ +70 °C
a
–40 °C ≤ T
≤ +65 °C
a
1)
–40 °C ≤ T
≤ +70 °C
a
1)
–40 °C ≤ T
≤ +65 °C
a